Gataca iLAS TIRF


Purpose:
- TIRF/ HiLo/ WF illumination
- Single-molecule imaging / tracking
- Super-resolution localization microscopy
- FRAP

Illumination / Imaging configuration: TIRF, HiLo / Widefield
Imaging contrasts: Fluorescence, phase contrast and transmitted light
Microscope stand: Inverted Nikon Ti2-E, motorized
Fluorescence excitation source: CW lasers (GATACA: https://www.gataca-systems.com/products/gataca-products/ilaunch/). 642nm@110mW; 568nm@140mW; 490nm@150mW; 410nm@100mW
Detection:
Fluorescence: 1 or 2 channels (Optosplit with ByPass mode)
- Brightfield / phase contrast: 1 channel
Detection device: ORCA-Flash 4.0

Camera Andor iXon Ultra 897 Retiga R1
Technology EM-CCD CCD
Detection

Fluorescence

Phase contrast, Brightfield

Chip size

512x512 pixels

1360x1024 pixels

Pixel size

16x16 µm

6.45x6.45 µm

Dynamic range

16-bit

14-bit

Max frame rate*

50 frames.s-1

25 frames.s-1

Quantum efficiency

95% @ 550 nm

75% @ 600 nm

*full chip, no binning

Acquisition software: Metamorph
Additional features:
XY motorized stage
- Hardware Z-drift compensation (Nikon PFS)
- Incubation box for temperature control
- External phase contrast
- Stroboscopic illumination for single-molecule tracking

Objectives:

Objective type Mag/NA Medium Contrast WB (mm) Z-Drift compensation system Pixel size (nm)

planApoϒ

20x / 0.75

Air

DIC N2

1

YES

400

planFlour

40x / 0.75

Air

DIC N2

0.66

/

660

Apo TIRF

100x / 1.49

Oil

Fluo

0.12

YES

125


Fluorescence filters single-channel detection:

Excitation dichroic filter

Emission filter

Reference

FPbase spectra viewer

Di03-R405/488/561/635-t1

525/50

FF03-525/50

link

Di03-R561

568 LP

BLP01-568R

link

Di03-R405/488/561/635-t1

609/54

FF01-609/54

link

Di03-R405/488/561/635-t1

708/75

FF01-708/75

link
 


Fluorescence filters split-channel detection:

Emission filter Reference

FPbase spectra viewer

 FF01-525/50

link

FF01-630/92

link

FF01-600/52

link

ET705/72

link